Within the Precautions and Warning Section, Polymorphism heading, of the Product's Package Insert, it stated that mutations at positions 20207, 20209, 20218 and 20221 exist and these mutations are spanned by the mutation probe. These rare mutations will lead to an unknown result after performing genotyping.
Source facts
Action
Product Advisory Notices were sent via e-mail on 3/12/09 to Roche Local Safety Officers and Roche county General Manager. Local Safety Officers and General Managers were to communicate to the end users with the Important Product Notice. In addition, Reagent Bulletin 09-052, dated 3/27/09, was issued to Roche U.S. Affiliate of U.S. customers.
Initiated
2009-03-12
Root cause
Labeling design
Terminated
2010-06-09
Distribution
Nationwide Distribution
Recall number
Z1006-2010
Recall status
Terminated
CDRH record ID
86392
FDA classified
2010-03-05
Recall event number
53708